National (State Primary) Standard of the Unit of Frequency Deviation of Frequency-Modulated Oscillations

Transferring of the size of frequency deviation unit to secondary and working standards by means of comparison with the help of comparator in accordance with DSTU 3392-96 (GOST 8.232-97) “Metrology. National verification scheme for measuring instruments of frequency deviation”.

The range of values of the frequency deviation ((f), which reproduces, stores and transmits the standard:

- from 10 Hz to 1 MHz - the main range;

- from 1 MHz to 8 MHz - extended range.

Modulation frequency range F:

- from 20 Hz to 200 kHz (main range);

- from 200 Hz to 500 kHz (extended range).

Carrier frequency range: 10 MHz, 50 MHz, up to 6 GHz.

Metrological characteristics

Range

From 10 GHz  to 1 MHz

from 1 MHz to 8 MHz

Uncertainty by type А (uA

3⋅10-4 Δf

(3–5)⋅10-4 Δf

Uncertainty by type В (uB)

(0,2–0,4)⋅10-3 Δf

(0,54–0,97)⋅10-3 Δf

(1,1 – 1,9)⋅10-3 Δf

Total standard uncertainty (uC)

5⋅10-4 Δf

(5–12)⋅10-4 Δf

Extended uncertainty (U)

1⋅10-3 Δf

(k = 2, p = 0,95).

(1–2,5)⋅10-3 Δf

(k = 2, p = 0,95).

Equipment that is part of the standard:

- frequency meter R&S HM8123,

- Keysight 336611A signal generator,

- vector generator Keysight N5172B,

- Keysight MSOS204 oscilloscope

- R&S FSV4 spectrum analyzer