State Primary Standard of the Unit of Harmonic Coefficient
Reproduction and maintenance of total harmonic distortion unit.
Transferring of total harmonic distortion unit to the secondary standard and working standards by means of the method of direct measurements and to the working measuring instruments by means of the method of direct comparison in accordance with DSTU 3391-96 (GOST 8.110-97) ”Metrology. National verification scheme for measuring instruments of total harmonic distortion”.
Metrological characteristics | |||||
Range | 0,01 kHz | 0,02 kHz | 0,2 kHz; 2 kHz; 20 kHz | 100 kHz | 200 kHz |
Uncertainty by type А (uA) | 0,0003·КГ | 0,0003·КГ | 0,0003·КГ | 0,0003·КГ | 0,0003·КГ
|
Uncertainty by type В (uB) | 0,0016+0,0052·КГ | 0,0005+0,0026·КГ | 0,0002+0,0016·КГ | 0,0005+0,0031·КГ | 0,0016+0,0031·КГ
|
Total standard uncertainty (uC) | 0,0016+0,0052·КГ | 0,0005+0,0026·КГ | 0,0002+0,0016·КГ
| 0,0005+0,0031·КГ | 0,0016+0,0031·КГ |
Extended uncertainty (U) | 0,0031+0,011·КГ | 0,0010+0,0052·КГ | 0,0003+0,0031·КГ
| 0,0010+0,0063·КГ
| 0,0031+0,0063·КГ |