State Primary Standard of Units of Average Power and Energy of Laser Radiation of Large Levels
Reproduction and maintenance of the mean power laser radiation units and high-class laser radiation energy with the wavelength of 10,6 μm. Transferring of the size of the mean power laser radiation units and pulse laser radiation energy to the working standards and directly to working measuring instruments by the method of direct measurements in accordance with DSTU “Metrology. National verification scheme for measuring instruments of power and highclass laser radiation energy”.
Metrological characteristics | ||
Range | from 10 J to 1000 W from 10 J to 1000 J | |
Uncertainty by type A (uA) | 0,5⋅10-2 | |
Uncertainty by type B (uB) | 0,41⋅10-2 | |
Total standard uncertainty (uC) | 0,65⋅10-2 | |
Extended uncertainty (U) | 1,3⋅10-2 |